°MEQU °M Station

The °MEQU °M Station enables medical providers to easily transition power supply for the °M Warmer utilizing Power Pack + to the °M-Station AC Power Supply. °M-Station provides continuous flow of warm fluids up to 9 liters per hour for use in-hospital and in road ambulances. The °MEQU Hospital Starter Kit System is indicated for use to warm blood, blood products, colloids and crystalloid solutions prior to parenteral administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.

Description

Light-weight, °M Station can be carried and easily operated by trained medical personnel. Warm blood / IV fluids from storage temperature (2°C to 6°C) to infusion temperature of 38°C ± 2°C in no more than ten (10) seconds. °M-Warmer System common plugs and ports enable rapid transitions from Power Pack or Power Pack + to °M Station. Permit set-up in no more than thirty (30) seconds. Unidirectional plug-in port prevents human error. Max Flow Rate Input Temp. of 5°C, Output Temp. min. 36°C 150 ml/min. Max Fluid Warming Volume: 9 Liters/Hr. Degree of protection against harmful ingress of Subjects / Water: °M-Station IP33. Electrical Compliance: EN 60601-1

Box Dimensions: 11”L x 6.5”W x 3”D
Weight: 3.63-Lbs.